Product Description
The unit eliminates powder contamination to protect the operator and the surrounding environment. It is designed to handle powder under aseptic conditions for optional G.M.P.
Technical Specifications
- Air cleanliness : ISO Class 5 / Class 100 / M 3.5 AREA
- Air Velocities : 0.45 ï 0.05 mps (90 ï 20 FPM)
- Air Flow : Vertical
- Noise Level : Less than 65 dB
- Vibration Level : Minimum
- POWER SUPPLY : 230V AC, 1-ï, 50HZ
- MOC : GI Powder Coated / SS-304 / SS-316
- Inside : Side panel with all three side with coving
Three Stage filtration
- EU4 " Pre-filters (90% down to 10 micron)
- EU7 - Intermediate Filter (95% down to 3 micron)
- EU14 - HEPA Filter - Supply / Exhaust (99.999% down to 0.3 micron)
- Statically & Dynamically balanced Motor-blower with isolator to reduce noise level
- Fluorescent Lights with Milky White Diffuser
- Differential Pressure Gauges :02 Nos. (Intermediate & HEPA Filter)
- ON/OFF Switches (BL / FL / UV)
- Auxiliary Powder Socket : 5/15 Amp for external equipment
- DOP Port
- DQ, IQ, OQ, PQ Documentation
Options and Accessories
A wide range of configurations to suit any containment application:
- Double skinned PUF Insulated side panels with/without view windows
- Stainless steel perforated grille for HEPA Filter
- Front Flexible Strip PVC Curtains
- SS Perforated Platform for keeping Weighing Scale (Fix /Moveable)
- Velocity display with High / Low Alarms
- Filter chocking alarms